Replacing Syringes, Tubing, and Valves in your C-Trap
Abstract
This guide provides step-by-step instructions for replacing syringes, tubing, and valves for the u-Flux fluidics module of your C-Trap or m-Trap. Procedures include material handling, tubing specifications, assembly and disassembly of tubing-to-valve connections, valve replacement, syringe maintenance, and leak testing with troubleshooting. Following these instructions helps maintain cleanliness, ensure correct installation of fluidic components, and minimizes the risk of leaks during operation. Keywords: fluidics maintenance, leak detection, fluidics troubleshooting
1. General Guidelines
1.1 Always wear clean gloves or sanitize them with ethanol before handling syringes, tubing, or valves.
1.2 Ensure the system contains MilliQ water only and no harmful substances before starting.
1.3 Remove valves from the Autofluidics unit before replacing tubing. It is not always necessary but strongly recommended for easier manipulation.
1.4 Handle threaded connections carefully and avoid over-tightening.
1.5 Replace tubing and syringes after 6–12 months of use to maintain optimal performance (C-Trap connection kit).
2. Materials
- Tubing material: FEP (Fluorinated Ethylene Propylene). For compatibility, refer to FEP chemical resistance charts.
- Available inner diameters for the C-Trap u-Flux module:
- Small: 0.1 mm (0.004") – Channels 1–5, for precious samples or low flow rates.
- Medium: 0.25 mm (0.01") – Typically Channel 6; can also be used for Channels 1–5 if higher flow is needed.
- Large: 0.75 mm (0.03") – For Channel 6 (e.g., waste) when Channels 1–5 use medium tubing.
- Syringes: Terumo Luer Lock, 3 mL (remove plunger before use).
3. Connecting tubing to valves

- Cut the SMALL size tubing with a tube cutter. Ensure a clean, perpendicular cut.
- Insert tubing through the beigetubing connection nut.
- Place a bluetubing connection Ferrule on the tubing with the wide part oriented toward the cut end.

- Ensure the tubing end is flush with the ferrule surface (not protruding or recessed), otherwise liquid might get caught there.
- Insert the assembly into the valve entry, keeping the tubing alignment intact (cut end in plane with blue inlet surface).

- Tighten the beigetubing connection nut fully, until it cannot be turned anymore. This should secure the tubing and ferrule without rotation.
- Once assembled, the tubing, the beigetubing connection nut, and the bluetubing connection Ferrule screw remain fixed together when unscrewed.
4. Disconnecting Tubing from Valves
4.1 Unscrew the beige tubing connection nut. The tubing and ferrule will rotate together during removal.
4.2 To separate, insert a fingernail or tool between the ferrule and screw, then slide the ferrule relative to the screw.
4.3 Replace tubing or ferrule as needed and reassemble, following Section 3.

5. Valve Removal and Attachment
5.1 Detaching a valve
- In Bluelake or the u-Flux applet, close all valves and open only the valve to be removed.
- Release fluidic pressure.
- Unscrew the valve mounting screws.
- Remove the valve.
5.2 Attaching a Valve
- In Bluelake or the u-Flux applet, close all valves and open the valve to be attached.
- Release fluidic pressure.
- Verify that the valve channel is in its open position before reinstallation.
- Position the valve and secure it with mounting screws.
6. Syringe Handling
6.1 Terumo 3 mL Luer Lock syringes can be reused if cleaned properly. Replace syringes that are stained, heavily soiled, or damaged.
6.2 To replace, unscrew the syringe from the transparent female Luer connector. In legacy systems, syringes may connect via blue manual valves.
6.3 Store plungers separately in a clean container for manual unclogging when required. Ensure that both the specific channel valve and the outlet valve are open during unclogging.
7. Leak Testing Procedure
7.1 Fill all syringes with water or another safe liquid, ensuring equal levels.
7.2 Flush all channels simultaneously.
7.3 Monitor fluid levels across all syringes; levels should decrease uniformly during flushing. Uneven drops indicate potential leaks.
8. Troubleshooting Fluidic Leaks
When performing leak testing, observe carefully where liquid emerges. Use the guide below to identify the source and apply corrective action. Always retest after implementing a step before moving to the next.
8.1. General Notes on Troubleshooting
- Always retest under low pressure before applying full operating pressure.
- Replace consumables (tubing, ferrules, syringes) if leak source cannot be reliably eliminated by reseating or tightening.
- If multiple leak points appear simultaneously, re-evaluate the assembly procedure (Section 3).
- For persistent leaks not resolved by the above steps, contact support@lumicks.com with system details and photographs of the affected area.